Vertical pulsating vacuum pressure steam sterilizer working principle:
Vertical Pulsating Vacuum Pressure Steam Sterilizer (Class B) working principle is generally: to take three mechanical vacuum forced discharge of cold air inside the sterilization chamber, to ensure that the sterilization chamber to reach saturated steam, the use of saturated water vapor effect on the load of microorganisms for a certain period of time, so that the microorganisms of the denaturation of proteins and thus lead to the death of microorganisms, in order to achieve the purpose of sterilization.
Application scenarios and sterilization test methods
It is suitable for medical and health care (including CDC), colleges and universities, food, drugs, cosmetics, biology, agriculture and forestry research and other units.
1, the product is suitable for all medical devices that can be sterilized by moist heat (fabric wrapping, liquid load, rubber load, solid load) sterilization. It is not recommended to be used in P3, P4 biological laboratory.
2、When the surgery is urgent to use the sterilization package, the supply center can not be supplied in time (the sterilization supply center is timed quantitative sterilization);
3、When liquid (culture medium) load sterilization is urgently needed;
4, when the fabric parcels (dressings, etc.) need to be sterilized (must use the pulsating vacuum pressure steam sterilizer)
5, when the sterilization preparation needs to dry (fabric packages, solid loads, glassware, etc.)
6, when there is a cavity load instruments need to be sterilized (must use pulsating vacuum pressure steam sterilizer)
7, especially suitable for primary health care institutions, medical equipment (fabric parcels, etc., need to dry preparations for sterilization.
Liquid sterilization test methods specified in YY/T1007-2018
5 x 1000ml beakers and temperature sensor test fixing position
Large volume sterilization chamber, each basket can put 4~5 1000ml beakers, according to user requirements can be equipped with 2~3 baskets.